IN-CHECK Introduction

What does it Look Like?

Which Inhalers?

How to Use

Accuracy

Inspiratory flow rates

Care and Cleaning

Validation data

Further Information

 

ACCURACY STATEMENT

 

Every In-Check DIAL is individually calibrated to ensure a high degree of accuracy. Inspiratory flows in the range of 15 l/min to 120 l/min can be measured. In-Check DIAL is accurate to +/- 10% or 10 l/min (whichever is the greater), and has repeatability of +/- 5 l/min.

In the In-Check DIAL has been tested by AEA Technology www.aeat.co.uk for accuracy, and has been verified as providing similar resistances to inhaler devices.

The In-Check DIAL has been calibrated using an ATS pulmonary waveform generator. In-Check DIAL complies with the AUS/NZ standard for back pressure in flow meters.

 

 

 

 

 

 

 

 

 

 

 

 

 


© by Clement Clarke International Limited 2000